Chapter 5 Multi-Unit Commands
Calibrating the Libraries in a Multi-Unit Configuration
Quantum P4000 and P7000 Libraries User’s Guide
89
Calibrating the Libraries in a Multi-Unit Configuration
5
The
Calibrate All PTMs
command enables you to calibrate all pass through
mechanisms in all libraries within the multi-unit configuration. Calibrate
the libraries during the initial installation and after any maintenance
procedure.
To calibrate all library PTM’s:
1
From the Master library in the multi-unit configuration, press the
Multi-Unit
tab.
The GUI displays the
Multi-Unit
screen (see
).
2
Press the
Calibrate All PTMs
button.
The GUI displays a
Command In Progress
dialog box while the
calibration process is in progress.
Figure 52 Multi-Unit
Screen
Note:
All libraries must be in standby mode.
Note:
Multi-Unit calibration MUST be executed from the master
library designated during multi-unit configuration.
Multi-Unit
System
Off-line
Standby
Load Port
Operator
Service
Multi
Unit
Back
Forward
Home
O
Configure
Multi-Unit
Calibrate
All PTMs
Stop
About