X-Ray Live Image Screen
After five seconds of holding the X-ray trigger, the system will begin emitting X-Rays, and transition to the
X-Ray Live Image
Screen
(Figure 7;
Settings
menu is hidden by default). System Status Indicators remain amber-colored, and the buzzer
pulse rate changes to two-second intervals.
Figure 7 X-Ray Live Image Screen
To change settings while emitting X-Rays, hold down the hat switch for three seconds (Figure 3) to show and activate the
Settings
menu (Figure 8).
Figure 8 Settings Menu Re-activated
The
X-Ray Live Image Screen
remains active until the red trigger button is released or 90 seconds has elapsed. When
either occurs, the following will happen:
System stops emitting X-rays
Buzzer turns off