8.6 How to use the Internal Flow (Optional)
The internal flow is an optional feature and must be specified at the time of manufacture
or the analyzer can be upgraded at service.
The GA5000 gas analyzer has the capability to measure the gas flow from a borehole
without the need for an additional flow pod.
This function can be selected as ‘on’ or ‘off’ for each ID using Landtec Systems Gas
Analyzer Manager. If it is off the user will not be prompted to measure the flow. However,
if LSGAM is not being used, flow measurement is automatically measured.
If the analyzer is fitted with the internal flow feature the internal flow option works on a
principle of a pressure drop across a known restrictor.
Instructions for Use:
1)
From the ‘Main Gas Read’ screen select the center soft-key for the ‘Special Actions’
command, this will open a sub-menu. Select Option number “4” for Flow.
2)
Making sure the sample tube is not connected to the blue port, select soft-key ‘Zero
Flow’.
3)
Move the sample tube from the white port to the blue port. The yellow exhaust hose
can be vented a safe distance from the sample point; do NOT re-circulate back into
the system.
Internal flow
4)
Select soft-key ‘Flow Options’ to toggle through and select type:
Average
Peak
Snapshot
Record
The average of the readings taken over the duration.
The highest value recorded over the duration.
The reading recorded at the point at which the ‘Store’ key is
pressed.
Multiple flow measurements recorded over the duration at a
user changeable interval (seconds).
Duration
The overall time that an average or peak reading can be
taken. Enter the value in minutes, or select continuous.