GA5000 Gas Analyzer
OMGA5KN4.6
Page
44
of
73
8.3 Answering ID questions
Prior to, or after, you have taken the reading, the ID questions should be populated, this is
only necessary when using ID questions, if not, please proceed with the reading as
normal. When you have uploaded IDs with ID questions assigned to them (LSGAM
software required), you will be prompted to answer these questions at the end of the
reading, this means that the answer only equates to that reading.
8.4 Special action
This menu enables the operator to perform the additional following functions out of
sequence if so desired.
1)
From the ‘Main Gas Read Screen’ select the soft-key ‘Special Action’ and the
following menu is displayed:
Special action menu
Note:
The list of special action options displayed on the special action menu is
dependent upon device type and sequence.
The following special actions may be performed:
Action
Function
Key 1 – Simple Gas
This action enables the operator to take a quick gas
reading. The pump will start running automatically
when this key is selected. The operator can stop the
pump by pressing the pump key on the keypad at
any time and the reading can then be stored by
selecting soft-key ‘Store’.
Key 2 – Site Questions
This action enables the operator to update site
questions prior to taking a reading.