5. To return to normal AUTO measurement mode, hold
down the MAX/MIN button for 2 seconds.
Peak Hold button
The Peak Hold function captures the peak AC or DC
voltage or current. The meter can capture negative
or positive peaks as fast as 1 millisecond in duration.
1. Turn the function switch to the A or V position.
2. Use the
MODE
button to select AC or DC.
3. Allow time for the display to stabilize.
4. Press and Hold the
PEAK
button until “
CAL
”
appears in the display. This procedure will zero
the range selected.
5. Press the
PEAK
button,
Pmax
will display.
6. The display will update each time a higher
positive peak occurs.
7. Press the
PEAK
button again,
Pmin
will display.
The display will now update and indicate the
lowest negative peak.
8. To return to normal operation, press and hold
the
PEAK
button until the
Pmin
or
Pmax
indicator switches off.