20
For inside use, max height:
Operating:3000m, Storage
10,000m
Pollution degree:
2
Safety:
The instrument complies with IEC 61010-1:2001
and IEC, 61010-031:2002
Power:
One 9V battery , NEDA 1604, IEC 6F22.
Dimensions:
182 (H) x 82 (W) x55 (D) mm
Weight:
Approx.: 375g.
Accuracy
Accuracy is given at 18 oC to 28 oC (65 oF to 83 oF), less
than 70 % RH
DC Voltage
(Auto-ranging)
Range
Resolution
Accuracy
400.0mV 0.1mV
+0.5% of rdg + 2 digits
4.000V
1mV
40.00V
10mV
400.0V
100mV
1000V
1V
+0.8% of rdg + 2 digits
Input Impedance: 7.8MΩ.
Maximum Input: 1000V dc or 1000V ac rms.