Document No.: 1P0446-1 Rev. B, April 7, 2014
14
Section 4 -
Installation
This manual outlines the requirements for system installation, work area allowance, and initial startup
procedure. The installer should refer to
warnings and
pages
before beginning installation.
4.1
Facilities Requirements
The specifications and requirements for the system and applicable options are listed as follows:
System
:
x
Weight
: 75 lbs. Designed for use on table top or counter
x
Dimensions (door closed)
: 13.75" W x 13.5" H x 21.5" L (349.25 x 342.9 x 546.1mm)
Chamber
:
x
Dimensions (interior)
: 5 x 5 x 6" Deep (127 x 127 x 152.4 mm)
x
Material
: 6061-T6 high grade aerospace aluminum
Electrical
:
110-220VAC, 1 phase, 10 amp, 50-60 Hz, 18 AWG, 3 wire. The power cord is supplied by PVA
TePla America Inc.
Process Gases
:
x
Regulated to 10 PSIG before entering the system
x
Connections made by
.25” O.D.
corrosion resistant materials such as stainless steel or Teflon
tubing (supplied by purchaser). Other materials can generate particulate matter via corrosion
which will clog gas shutoff valves.
Gas Fittings
:
x
1/4" FPT fittings (the gas lines will need
¼” Swagel
ok ferrule style compression fittings--to
connect the incoming gas to the back of the IoN 3 MHz system)
Vacuum Pump
:
x
a KF25 flange is required to connect the vacuum hose to the IoN 3 MHz.
Internal RF Power Generator:
x
150 Watts +/-5%@ full load output into 50 Ohm impedance
x
13.5MHz operating frequency
x
Solid state circuitry
x
Automatic impedance matching