Follow the procedure below to help avoid any problems.
Enter the service code (552621) and change the following parameters.
If you have not done so already choose Solids in P102.
This will also change the following parameters.
P71 = Update Datem = 0.
Using the masking technology require valuable
processing time, and is not necessary for these types of applications.
P88 = Fixed Distance
(Gate Width) = 0.3m
Go to Parameter P882 and chose Fast Response = 1
This will also change the following parameters.
P44 = Breakpoint = 90.
Changing this parameter will ensure no gain is applied
to the echo saving processing time.
P45 = Averages = 1.
Changing this parameter will decrease the number of
averages taken before the echo is processed thus saving processing time.
Now Change the following parameters these are not adjusted by choosing Solids
in P102 or Fast Response in P882.
P67 = Echo Algorithm
= 2 = Largest. The unit will measure on the largest echo
seen above the loss limit Threshhold.
P79 = Use Gate = 0.
Not using the gate means that the unit will measure on the
largest echo seen above the loss limit Threshhold that is within the entire length
of the range. Ensuring that the level can be followed it if moves very quickly, and
might normally move outside the measuring window.
P870 = Fill Damping = 99999.
Setting the empty and fill damping parameters to
the maximum value will mean the unit effectively has no measurement damping.
P871 = Empty Damping = 99999.
See above.
P20 enter 2 (Default Datem)
This must be the last thing you do after entering
the above parameters.
Summary of Contents for ULTRA 5
Page 1: ...ULTRA Series SERVICE MANUAL ...
Page 2: ......
Page 4: ......
Page 8: ...Page 2 This page left blank intentionally ...
Page 12: ...Page 6 This page left blank intentionally ...
Page 18: ...Page 12 This page left blank intentionally ...
Page 24: ...Page 18 This page left blank intentionally ...