P044 Slope
Angle at which the rate of signal amplitude decay compensation (gain) is applied
over distance starting at P43 (see table above).
Values Min = 0
Max = 100
P045 Num Averages
Number of processing cycles or transducer ‘shots’ that are averaged prior to
processing of the echo trace.
Values Min = 1
Max = 50
P047 Outside Count
This parameter determines the number of processing cycles, or transducer ‘shots’,
that any valid echo must be seen outside of the gate before the gate will move and
allow the new echo to be processed. This will prevent any unwanted update, or
false readings being given due to the intermittent appearance of echoes from such
devices as stirrers.
Values Min = 1
Max = 5000
Default = Transducer Dependant
Default = 3
Default = 10