Debug Level
P086 Ping Delay (From Software Revision 6.4.1)
This parameter is used to delay the time between each ping in order to prevent the
risk of a second ping interfering with the initial return echo and causing
problematic readings. It delays the time between each ping.
P087 Fail Time 2
This parameter determines the time after an event, such as “Loss of Echo”, that a
warning message will appear in the Auxiliary line of the display.
Values: Min. 0.1 min
Max. 9999 min
P088 Hard Default
P88 allows both User and Service parameters to be reset to their Factory Default
values.
N.B. P930 Factory Defaults will allow the user to reset only those parameters
which are accessible via the User Passcode (Default 1997), but will not reset any
parameters which are accessed via the Service Code 552621. Thus ensuring any
adjustments carried out by Service personnel are protected.
Options:
0 = No
1 = Yes
Default = 1 min
Default = 0