MAIN DISPLAY
The main display shows the units selected from the
Units/Mode
menu,
Flow or Velocity rate being measured, TOTALIZER and RELAY states. The
AVFM 6.1 will start-up with this display.
MESSAGE ICON
Press
from the main display to view temperature measurement, status of
the data logger and error/warning messages provided by the instrument. The
Message Icon will appear on the main display if error messages are being
generated by the instrument. Press
to return to the main display.
STATUS
Press
from the MAIN display to view instrument status.
Level
Displays the measured level in units selected in
the Units/Mode menu.
Velocity
Displays the measured velocity in units selected
in the Units/Mode menu.
Flow
Displays the flow rate in units selected in the
Units/Mode menu. The flow is calculated based
on the individual level and velocity data, and
the channel shape and size programmed in the
Channel Setup menu.
Min Flow
Displays a read-only value of the minimum
flow cutoff. Any flow rate measured below this
Min Flow will be displayed as 0 on the LCD
display.
Signal Strength
Displays the strength of the received Doppler
velocity signal on a 0-100% scale.
Signal Cutoff
Displays a read-only value for signal cutoff. If
the Signal Strength is less than the Signal
Cutoff, velocity will be reported as 0. This
setting may need to be adjusted in case of
unstable velocity measurements with no water
moving, or when high levels of industrial noise
are present. The Signal Cutoff can be adjusted
in the Calibration menu.
USG/m
Tot 20130.8 USG
Relays 1 2 3 4 5 6
0.000
--Message-----------
Data Log
Logging
d
Log Use
0%
Sensor
Good
Temperature
24C
--Status------------
Level
0.00 ft
Velocity
0.00ft/s
Flow
0.000 ft/s
Min Flow
0.000USG
Signal Strength
0%
Signal Cutoff
5%
Echo Confidence 100%