PULSAR MEASUREMENT
29
INSTALLATION - (cont.)
1.
Flow Profile Distortion
The pipe/channel in which the QZ02 sensor is mounted should be free of bends, tees, sudden changes in
slope, and there should not be objects in the pipe/channel which disturb the flow profile in front of the
sensor.
In general, the QZ02 sensor should be mounted with at least 10 pipe diameters or channel widths of
straight-run upstream, and 5 pipe diameters or channel widths downstream: