FLOWCERT LITE (UL) INSTRUCTION MANUAL
20
Cable Entry
There are 6 cable gland knockouts on the base of the FlowCERT Lite (5 x
20mm (0.79"), 1 x 16mm (0.63")) and 4 on the rear (4 x 18mm dia (0.73")).
Select which ones you wish to take out, and remove them by using a circular
cutter, such as a tank cutter. Take care not to damage the circuit board
inside whilst undertaking this. Do not use a hammer, as this may cause
damage to the enclosure. It is recommended that you use suitable cable
glands to ensure that the ingress rating is maintained.
Terminal connection details
There is a wiring diagram inside the terminal access cover, and is shown
below:
Terminal connections
Power
The FlowCERT can operate from mains AC and automatically from DC or
battery backup in the event of power failure or can be operated permanently
from DC or batteries.