![Pulsar Measurement FlowCERT Instruction Manual Download Page 12](http://html1.mh-extra.com/html/pulsar-measurement/flowcert/flowcert_instruction-manual_1625413012.webp)
FLOWCERT LITE (UL) INSTRUCTION MANUAL
12
Product Specification
PHYSICAL
Wall Mount Outside Dimensions
9.25 x 7.24 x 4.72 inch (235 x 184 x 120 mm)
Weight
Nominal 2.2lbs (1kg)
Enclosure Material/Description
Polycarbonate, flame resistant to UL94-5V
Cable Entry Detail
10 cable entry knock outs, 1 x M16, 5 x M20
underside, 4 x 18mm dia (PG11) at rear
Transducer Cable Extensions
3-conductor 20AWG screened
Maximum Separation
3,250 ft (1,000m), 1,640 ft (500m) for mmWave
ENVIRONMENTAL
IP Rating (Wall)
IP65 (NEMA 4X), IP20 when open
Max. & min. temperature
(electronics)
-20 ºC to +50 ºC (-
4˚F to +120˚F
)
Flammable atmosphere approval
Safe area: compatible with approved dB transducers
(see transducer spec' sheet)
APPROVALS
CE Approval
See EU Declaration of Conformity
UL
Certificate Number E257330
PERFORMANCE
Accuracy
0.1%
of the measured range (dB3 and external
temperature compensation or internal
compensation using twin sunshields)
Resolution
0.5mm (dB3 and external temperature compensation)
Max. Range
Dependent on application and transducer (maximum
52.49ft (16m) dBR16
Min. Range
Dependent upon application and transducer
(minimum zero dB Mach3)
Rate Response
Fully adjustable
ECHO PROCESSING
Description
DATEM (
D
igital
A
daptive
T
racking of
E
cho
M
ovement)