FLOWCERT LITE (UL) INSTRUCTION MANUAL
106
P717 Exponent
This parameter is used to enter the exponent value when:
P700 PMD Type =1 Exponent and P701 Primary MD = 7 others.
P718 K Factor
This parameter is used to enter the K Factor when: P700 PMD Type = 1
Exponent and P702 Calculation = 1 Absolute
see table below for further
details
.
PRIMARY MEASURING DEVICE
P718 K FACTOR
P700 = 1 Exponent
P701 = Supp. Rectangular Weir
Automatically calculated
P700 = 1 Exponent
P701 = 2 Trapezoidal
Automatically calculated
P700 = 1 Exponent
P701 = 3 Venturi Flume
Obtain value and enter
P700 = 1 Exponent
P701 = 4 Parshall Flume
Automatically calculated
P700 = 1 Exponent
P701 = Leopold Lagco Flume
Automatically calculated
P700 = 1 Exponent
P701 = 6 V-Notch
Automatically calculated
P700 = 1 Exponent
P701 = 7 Other
Obtain value and enter
P719 Throat Width
This parameter is used to select the Throat Width of the flume when: P700
PMD Type = 1 Exponent and P701 = 4 Parshall Flume. After selecting the
Throat Width, the Exponent P717 and K Factor P 718 will be set
automatically.