4.
Press EXIT to start calibration.
ð
The number counts down from 30 in the lower display.
5.
At the end of the process, the lower display alternates
between 10 and the EXIT key and asks for the 10.0 NTU
standard solution.
A0525
Fig. 42: [10 NTU calibration] display
A0524
Fig. 43: [Start calibration] display
6.
If the alternating display of 10 and the EXIT key is not dis‐
played, press the DOWN key until 10 appears on the display.
7.
Insert the 10 NTU standard solution. Rotate the standard sol‐
ution in the measuring chamber so that the lowest value is
displayed in the upper line of the display.
8.
Press EXIT to start calibration.
ð
The number counts down from 60 in the lower display.
9.
At the end of the process, the lower display alternates
between 02 and the EXIT key and asks for the 0.02 NTU
standard solution.
Operation