4.1 Wall-mounted
NOTICE!
Distance to the sampling site
Install the DULCO turb C no more than 3 metres
away from the sampling site. This is the only way
to ensure a quick response time from the system.
Horizontal position
Ensure that your device is horizontal.
Clearance above the sensor
Leave a minimum clearance of 200 mm above the
DULCO turb C for any work required on the
DULCO turb C.
Fasten the DULCO turb C using four M6 screws,
and fix the terminal box in place using two M4
screws.
A0515
A
B
Fig. 1: The drawing is not to scale. All values are in mm.
Installation (mechanical)
Assembly and installation