ð
The DULCO turb C must be reset to enable the new des‐
sicant to be detected. To do this, disconnect the sensor
connection cable from the DULCO turb C for 2 seconds
and then reconnect the sensor connection cable. Other‐
wise, the warning
[DESC] can appear on the LCD dis‐
play.
A0648
Fig. 12: Insert in the packaging for the dessicant bag
Always use the original packaging. If you do not have any original
dessicant, use 113 ... 170 grammes of zeolite-based dessicant
(molecular sieve) of similar quality (3 Å pore width). Silica gel or
other chemical dessicants may not be used.
6.2 Starting the system
The following steps describe the measurement process with this
device:
1.
Switch on the device and allow it to warm up. 45 ... 60
minutes during initial commissioning.
2.
Check whether the cuvette is tight before starting the flow.
Hold the cuvette at the top 12 mm of it to prevent fingerprints
impairing the measured values.
Commissioning