VIII.A. Operational Mode Setup Requirements (continued)
(i) Setting the Maxwell
®
16 Instrument Operational Mode
(ii) Changing the Maxwell
®
16 Instrument Operational Mode
1. Make sure the instrument door is closed and turn on the Maxwell
®
16 Instrument. The screen will display
both the firmware version number and the current operational mode setting.
2. Verify that the current operational mode displayed matches the
operational mode required by both the installed hardware and the
purification kit to be used (see Table 2).
3. If the operational mode displayed does not match the operational
mode required by the purification kit, you will need to change the
instrument setting (see (ii) below, “Changing the Maxwell
®
16
Instrument Operational Mode”).
60
33
ME
Fnsc
Promega MAXWELL 16
Purification System
Version 4.0
LEV
Operational Mode Setting
Rsch = Research
Fnsc = Forensic
IVD = Clinical
SEV = Standard Elution Volume
LEV = Low Elution Volume
1. Press the “Menu” button to get to the Menu screen.
2. Scroll down to move the cursor to “Setup”. Press “Run/Stop” to
select “Setup”.
5
3
14MA
1.
2.
Demo
3.
Setup
Run
====Menu====
3. Scroll up/down to move the cursor to the desired operational
mode. Press “Run/Stop” to select.
Refer to Table 2 for information on the operational mode required
for research and forensic Maxwell
®
16 Purification Kits.
5840MB
1.Research Mode
2.Forensic Mode
3.Clinical
==Operational Setup==
4. Scroll down to move the cursor to the desired hardware mode.
If the instrument is configured with the Standard Elution Volume
hardware, select “SEV”. If the instrument is configured with the
Low Elution Volume hardware, select “LEV”.
Press “Run/Stop” to select.
60
3
5MA
1.SEV Hardware
2.LEV Hardware
==Hardware Setup==
!