4.
Gently lift up the SEV Magnetic Rod Assembly and
remove it.
5.
Insert the black LEV Plunger Bar Adaptor onto the
bottom of the remaining arm. Make sure that the side
containing the screws faces upward. Several magnets
will “click” and hold the LEV Plunger Bar Adaptor in
place once it is attached. The LEV Plunger bar should
then be fairly difficult to remove.
6.
Insert the black LEV Magnetic Rod Assembly. The
screws should be facing toward you. Press firmly to
place the LEV Magnetic Rod Assembly in position. It
will fit tightly into place.
6
3
91T
A
O
r
ie
n
tatio
n
of the Mag
n
etic Rod Assembly
front
back
7.
Secure the LEV Magnetic Rod Assembly with the three
thumbscrews from the standard magnetic rod
assembly, and hand-tighten. Complete the
configuration by updating the Firmware to the LEV
setting (Section VIII.A).
Failure to change the instrument firmware
operational mode from SEV to LEV will cause
damage to the instrument.
!
SEV Magnetic Rod Assembly
LEV Plunger Bar Adaptor
tm300.0208_Corrected.qxp 3/11/2008 10:42 AM Page 31