6962MB
Calibration Error!
Continue?
Error Code: 3
Yes No
After completion,
Refer to
Tech. Manual #TM300
for troubleshooting
Turn off instrument!
Refer to
Tech. Manual #TM300
for troubleshooting
Yes
No
Refer to Troubleshooting
Section for a detailed list
of error codes.
IV. Operating the Maxwell
®
16 Instrument
IV.A. Navigation
The system firmware will prompt you through initiation of a purification run. Follow the directions
displayed on the LCD screen. Use the “Scroll Up” and “Scroll Down” buttons to move the cursor to
the desired position. Once the cursor is moved to the desired position in the list, press the “Run/Stop”
button to make the selection. At any time during the selection process, you may press the “Menu”
button to return to the beginning.
IV.B.Operational Qualification
With instrument power on:
Each time you turn the instrument on, it will automatically go through a self-diagnostic test during
which the platform, plunger bar and magnetic rod assembly are moved to check that the instrument is
functioning properly. Upon completion of this test, the screen will display either “Diagnostic Check
Successful” or “Calibration Error”.
During instrument runs:
During runs, the instrument will automatically do periodic self-diagnostic tests to verify that the
platform, plunger bar and magnetic rod assembly are within calibration during the purification
procedure. If the instrument detects that any of these parts are not within calibration during a run,
the method will pause and display “Calibration Error”.
If calibration error occurs:
If a calibration error occurs, you will see the “Calibration Error” screen. Note the Error Code Number
displayed on the LCD screen, and refer to the Troubleshooting section of this manual (Section VI.C) for
more information on each Error Code. If you are in the middle of a run, you can choose to continue the
purification procedure or abandon the run by turning off the instrument. Continuing the run will not
damage the Maxwell
®
16 Instrument but may result in suboptimal recovery of purified nucleic acid.