30
TMBA_B-BAPS-e-2010
12 Instant help
In case of a fault in the program sequence, the balance should be shortly switched
off. The weighing process must then be restarted from the beginning.
Failure:
Possible causes:
The displayed weight
does not glow.
The balance is not switched on.
The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
Power supply interrupted.
The rechargeable battery / the battery is/ are
inserted incorrectly or empty
No rechargeable battery / no battery is/ are inserted
The displayed weight is
permanently changing
Draught/air movement
Table/floor vibrations
The weighing pan is in contact with foreign bodies
or is not correctly positioned.
Electromagnetic fields/ static charging (choose
different location/switch off interfering device if
possible).
The weighing result is
obviously incorrect
The display of the balance is not at zero.
Adjustment is no longer correct.
Great fluctuations in temperature.
The balance is on an uneven surface.
Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
No Data can be
transmitted on WIFI
interface
The network signal is not stable or too weak.
The Interface is not configured correctly.
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.