26
TMBA_B-BAPS-e-2010
9 Menu
Access to service menu “tCH” is locked in verified balances.
To disable the access lock, destroy the seal mark and actuate
the adjustment switch. For position of adjustment switch, see
chap. 13.
Attention
:
After destruction of the seal the weighing system must be re-
verified by an authorised agency and a new verification
wire/seal mark fitted before it can be reused for applications
subject to verification.
9.1 Navigation in the menu
Call up menu
Switch-on the balance, in weighing mode press
for
approx. 3 sec., until the first function
AF
appears.
Select function
With help of
or
, the individual functions can
be selected one after the other.
Change settings
Confirm selected function by
. The current setting
will be displayed.
Select desired setting with
or
and
acknowledge by
or reject with
.
Exit menu/
Return to
weighing mode
Press
,
Exit
will be displayed. Press
, the
balance returns into the weighing mode.