Operate
page 22
last update: 30.07.2014
For calibrating you need to choose
“2
nd
standa
rd”.
Afterwards you will be shown the information. The
standard value needs to be consistent with the
retroreflection values mentioned on the label which is
on the backside. Adjust the standard value if
necessary.
If one of the geometries (RL or Qd) is deactivated,
you can activate them by clicking on the line
„Calibrate“ and choose „On“.
Please check the calibration settings also for the Qd
geometry and if necessary change them accordingly.
Afterwards press the backwards arrow
for
getting back to the calibration menu.
Depending on the setup settings the deviation of the
2
nd
standard will be shown or hidden. By pressing
and selecting “Setup” and “Calibration”
you can change these settings. With
confirm the calibration and you get back to the
measure mode.