14
13
4. Maintenance
• Do not attempt any repairs to the device or any of its accessories.
Contact us at 1-800-363-2381 for repair information.
• The manufacturer will not be held responsible for the results
of maintenance or repairs by unauthorized persons.
• Check the device before each use for signs of wear and/or
damage. Replace worn items (electrodes, lead wires) as required.
3. Cleaning the electrode cords
Clean the electrode cords by wiping them with a damp cloth.
Coating them lightly with talcum powder will reduce tangles
and help them last longer.
5. To replace the battery
To ensure safe and adequate functionning of this device, it is
necessary to periodically change the battery. To do so, slide the belt
clip off, then the battery compartment cover. Remove the old
battery and insert a new one, matching the positive and negative
ends of the battery to the markings, then close the compartment.
Caution:
If a battery has leaked, avoid any contact with skin, eyes
and mucous membranes. Flush any
affected skin with water and contact
a medical professional right away.
Do not attempt to recharge
the battery supplied with this
device. If you plan not to use
the device for a long time,
remove the battery and store
appropriately, away from heat.
THERAPY PROGRAMS
Program
Time Rate
Width
Wave description
TENS
Burst
30 100Hz 200
µ
s 9 pulses/burst, 2 bursts/sec.
Normal
30 100Hz 200µs Continuous
Mod.
30 100Hz 200µs First 0.5 sec.: pulse width at 50%
setting. Next 0.5 sec.: pulse rate
at 50% setting and width at 100%
setting.
SD1
30 100Hz 120 ~ Set intensity goes from 100% to
200µs 60% then back to 100% while set
pulse goes from 60% to 100% and
back to 60%, over 10-second cycles.
SD2
30 100Hz
60 ~
Set intensity goes from 100% to
200µs 30% then back to 100% while set
pulse goes from 30% to 100% and
back to 30%, over 10-second cycles.
Shoulder 30
2 ~
60 ~
12-second cycles, span: 50%.
125Hz 200µs Rate stays in the 2-10Hz range for
1/3 of the cycle time as it
modulates down to 2Hz and back
up again.
Lower
30 100Hz
300/
1-second cycles, span: 60%.
back
200µs Modulated amplitude via pulse width
0.5 sec. at 100% of set intensity and
0.5 sec. at 50%.
Elbow
30
2 ~
100/
12-second cycles, span: 50%
125Hz 200µs Rate stays in the 2-10Hz range for
1/3 of the cycle time as it
modulates down to 2Hz and back
up again.
Knee
30
50Hz
200µs 6-second cycles (1 sec. up / 1 sec.
down) then 6 seconds rest.
EMS
A
30
50Hz
200µs 6-second cycles (1 sec. up / 1 sec.
down) then 6 seconds rest.
S
30
50Hz
200µs 8-second cycles(2 sec. up / 2 sec.
down) then 4 seconds rest.
D
30
40Hz
200µs 11-second cycles (4 sec. up / 2 sec.
down) then 15 sec. rest (1 sec. delay).