TROUBLESHOOTING

If your device does not seem to be operating correctly, refer to the 
chart below to determine what may be wrong. Should none of these 
measures correct the problem, the device should be serviced.

Problem

Possible cause

Solution

Display fails 
to light up

Battery contact failure

1) Install new batteries

2) Ensure that the battery is inserted 
    correctly, respecting the polarities, 
    and that battery contacts are in 
    place and in good condition

Weak or no
stimulation

Electrodes are dried 
out or dirty

Electrodes are 
improperly positioned

Old / worn / damaged 
lead wires

Replace and re-connect the electrodes

Replace the lead wires

Stimulation is
uncomfortable

Intensity is too high

Decrease intensity

Electrodes are too
close together

Damaged or worn
electrodes or lead wires

Reposition the electrodes

Replace the lead wires

Electrode active area 
is too small

Replace electrodes with ones that are 

2

16 cm  (4 cm x 4 cm or 1.5” x 1.5”) 
or larger

The user instructions are
not being followed

Please check the manual before use

Intermittent
output

Lead wires

1) Verify connection is secure 
     and firmly in place

2) Turn down the intensity. Rotate lead 
     wires in socket 90°. If the problem 
     persists, replace lead wire.

3) If replacing the lead wire does not 
    solve this problem, contact us at: 
    1-800-363-2381

Program option in use

Some programs will seem intermittent. 
This is expected. 

Reposition and re-connect 
the electrodes

THERAPY PROGRAMS (CONTINUED)

Program 

Time  Rate 

Width 

Wave description

EMS

  Neck 

30 

20Hz 

180µs  Continuous

 

Abdomen  30 

65Hz 

250µs  10-second cycles (2 sec. up / 2 sec.

 

 

 

 

 

down) then 8 seconds rest.

 

Hip 

30 

50Hz 

250µs  13-second cycles (1 sec. up / 2 sec.

 

 

 

 

 

down) then 10 sec. rest.

 

Leg 

30 

40Hz 

200µs  13-second cycles (1 sec. up / 2 sec.

 

 

 

 

 

down) then 10 sec. rest.

 

Foot 

30 

65Hz 

200µs  10-second cycles (2 sec. up / 2 sec.

 

 

 

 

 

down) then 8 seconds rest.

SKIN CARE RECOMMANDATIONS

• Always clean the treatment area using a mild soap, rince skin 

thoroughly and dry well before applying the electrodes and after 
removing them.

• Clip excess hair with scissors – do not shave the treatment area.

• Your doctor or therapist may recommend the use of a skin prep 

solution. Let it dry thoroughly before applying the electrodes. 

• Avoid stretching the electrode over the skin as this could cause 

"pulling stress". Apply the electrode from the center outward to 
prevent this from happening.

• When removing the electrodes, pull in the direction of hair growth.

• It may be helpful to apply a skin lotion over the treatment area when

not using the electrodes.

• Never apply the electrodes over irritated or broken skin.

Summary of Contents for alevia

Page 1: ...ev1 1012 P N 861 715425 Please read this manual thoroughly before using this device for the first time Physiotherapy Physiothérapie 2 in 1 Physiotherapy Device 715 425 TENS instructions rev1 1012 715 425 861 715425 24 811 cm x 18 81 cm Folded 10 5 cm x 15 cm Open 21 cm x 15 cm This is a double sided saddle stitched instruction booklet Isabelle Picard TM ...

Page 2: ...lace electrodes on skin Turn the device on Select the therapy program mode and site Set the pulse width and rate Set the treatment time Adjust the intensity for each channel and begin treatment After treatment is finished To consult the memory To turn off the device Low battery indicator CLEANING AND MAINTENANCE 12 Cleaning the device and accessories Maintenance To replace the battery THERAPY PROG...

Page 3: ...is can help improve and maintain muscle tone without actual physical activity and therefore prevent atrophy SAFETY INFORMATION 1 This device is recommended for Symptomatic relief of chronic intractable pain acute post traumatic pain or acute post surgical pain Improving local blood circulation Relief from muscle spasms 2 Do not use in the following situations Undiagnosed pain syndromes until etiol...

Page 4: ...administered Do not use it near highly flammable substances gases or explosives Apply the electrodes to clean dry and unbroken skin only Keep electrodes separate during treatment Electrodes coming in contact with each other could result in improper stimulation or skin burns 5 Possible adverse reactions Skin irritation or electrode burn under the electrodes Allergic skin reaction to electrode gel m...

Page 5: ...ctor into the electrodes connector ensuring that no bare metal remains exposed 3 Connect lead wires to the device This device has 2 output channels which can be used simultaneously with 2 electrode pairs or individually 1 channel with 1 electrode pair After making sure that the device is turned off insert the wire plugs into the sockets at the top of the unit Caution Always use the lead wires supp...

Page 6: ...n the device on Carefully read the contraindications and safety information in this manual before using for the first time Turn the intensity control for either channel 1 or 2 clockwise The device will beep Continue as described below before increasing output intensity 6 Select the therapy program mode and site Press M until the required mode EMS or TENS and body part are selected The selected the...

Page 7: ...lace It may be helpful between applications to moisten their adhesive surface with a few drops of water be careful not to over saturate and then let them air dry to help them last longer Reusable self adhesive electrodes 10 After treatment is finished Once the treatment time has elapsed turn the intensity control dials counter clockwise to the OFF position Unplug the lead wires from the device To ...

Page 8: ...move the battery and store appropriately away from heat THERAPY PROGRAMS Program Time Rate Width Wave description TENS Burst 30 100Hz 200µs 9 pulses burst 2 bursts sec Normal 30 100Hz 200µs Continuous Mod 30 100Hz 200µs First 0 5 sec pulse width at 50 setting Next 0 5 sec pulse rate at 50 setting and width at 100 setting SD1 30 100Hz 120 Set intensity goes from 100 to 200µs 60 then back to 100 whi...

Page 9: ...the problem persists replace lead wire 3 If replacing the lead wire does not solve this problem contact us at 1 800 363 2381 Program option in use Some programs will seem intermittent This is expected Reposition and re connect the electrodes THERAPY PROGRAMS CONTINUED Program Time Rate Width Wave description EMS Neck 30 20Hz 180µs Continuous Abdomen 30 65Hz 250µs 10 second cycles 2 sec up 2 sec do...

Page 10: ...C with a relative humidity of 10 90 atmospheric pressure from 700 hpa to 1060 hpa Dimensions 110 x 65 x 30 2 mm Weight 139 g with battery Problem Possiblecause Solution STORAGE Remove the battery if the device is not in use for long periods of time Store the device in a dry place and protect it against heat sunshine and moisture Never place any heavy objects on the device DISPOSAL Used fully disch...

Reviews: