LIMITED WARRANTY

AMG Medical Inc. warrants the stimulator to be free from defects in 
material and workmanship for a period of one (1) year, to be proven by 
means of the sales receipt or invoice.  This warranty is valid for the 
original purchaser only.  Any alterations, abuse, misuse or accidental 
damage voids this warranty. Repairs under warranty do not extend the 
warranty period. For service under warranty, call us at: 
1-800-363-2381, between 8 AM and 5 PM EST.
• The following is excluded under the warranty:
  A) All damage which has arisen due to improper treatment, 
    e.g. nonobservance of the user instructions.
  B) All damage which is due to repairs or tampering by the customer
    or unauthorized third parities.
  C) Damage which has arisen during transport from the manufacturer
    to the consumer or during transport to the service centre.
  D) Accessories which are subject to normal wear and tear.
• Liability for direct or indirect consequential losses caused by the 
  unit are excluded even if the damage to the unit is accepted as a
  warranty claim.

TECHNICAL INFORMATION

• Channel: Two isolated channels
• Power supply: 1 x 9V battery, type 6F22
• Waveform: Single square pulse
• Pulse duration: adjustable, from 30 to 350μs. 
• Pulse frequency: adjustable, from 1 to 150Hz
• Treatment time: 5 to 60min
• Amplitude: 0-40V(Vpp)(Large output 40V

±

10% at 500ohm load)

• Operating conditions: 0°C to 40°C with a relative humidity 
  of 30% - 75%, Atmospheric pressure from 700 hpa to 1060 hpa
• Storage conditions: -10°C to 55°C with a relative humidity 
  of 10%~90%, atmospheric pressure from 700 hpa to 1060 hpa
• Dimensions: 110 x 65 x 30.2 mm 
• Weight: 139 g (with battery)

Problem

Possible cause

Solution

STORAGE

• Remove the battery if the device is not in use for long periods 
  of time.
• Store the device in a dry place  and protect it against heat, 
  sunshine and moisture.
• Never place any heavy objects on the device.

DISPOSAL

Used fully discharged batteries must be disposed of in a specially 
labeled collection container, at toxic waste collection points or through 
an electrical retailer. 

Please dispose of the device in accordance with the directive 
2002/96/EC WEEE (Waste Electrical and Electronic Equipment). 
Contact your local distributor for information regarding disposal 
of the unit and accessories.

Stimulation 
is ineffective

Improper electrode 
placement

Reposition electrode

Unknown

Contact your doctor

The skin 
becomes 
red and / or 
you feel a 
stabbing pain

Repetitive use of 
electrodes on the 
same site

Reposition the electrodes. If at any 
time you feel pain or discomfort, stop 
use immediately

Electrodes aren’t 
applied on the skin properly

Ensure the electrodes are attached 
securely to the skin

The electrodes are dirty

Clean the electrode pads with a damp, 
lint-free cloth or replace with new 
electrodes. Clean the electrode cord 
according to the description in the 
user manual

The surface of the 
electrode is scratched

Replace with a new electrode

Output current
stops during 
therapy

The electrode pads 
come off the skin

Turn off the device and attach the 
electrode firmly to the skin

A cable is disconnected

Turn off the device and connect the cable

Batteries are low

Replace with new batteries

Summary of Contents for alevia

Page 1: ...ev1 1012 P N 861 715425 Please read this manual thoroughly before using this device for the first time Physiotherapy Physiothérapie 2 in 1 Physiotherapy Device 715 425 TENS instructions rev1 1012 715 425 861 715425 24 811 cm x 18 81 cm Folded 10 5 cm x 15 cm Open 21 cm x 15 cm This is a double sided saddle stitched instruction booklet Isabelle Picard TM ...

Page 2: ...lace electrodes on skin Turn the device on Select the therapy program mode and site Set the pulse width and rate Set the treatment time Adjust the intensity for each channel and begin treatment After treatment is finished To consult the memory To turn off the device Low battery indicator CLEANING AND MAINTENANCE 12 Cleaning the device and accessories Maintenance To replace the battery THERAPY PROG...

Page 3: ...is can help improve and maintain muscle tone without actual physical activity and therefore prevent atrophy SAFETY INFORMATION 1 This device is recommended for Symptomatic relief of chronic intractable pain acute post traumatic pain or acute post surgical pain Improving local blood circulation Relief from muscle spasms 2 Do not use in the following situations Undiagnosed pain syndromes until etiol...

Page 4: ...administered Do not use it near highly flammable substances gases or explosives Apply the electrodes to clean dry and unbroken skin only Keep electrodes separate during treatment Electrodes coming in contact with each other could result in improper stimulation or skin burns 5 Possible adverse reactions Skin irritation or electrode burn under the electrodes Allergic skin reaction to electrode gel m...

Page 5: ...ctor into the electrodes connector ensuring that no bare metal remains exposed 3 Connect lead wires to the device This device has 2 output channels which can be used simultaneously with 2 electrode pairs or individually 1 channel with 1 electrode pair After making sure that the device is turned off insert the wire plugs into the sockets at the top of the unit Caution Always use the lead wires supp...

Page 6: ...n the device on Carefully read the contraindications and safety information in this manual before using for the first time Turn the intensity control for either channel 1 or 2 clockwise The device will beep Continue as described below before increasing output intensity 6 Select the therapy program mode and site Press M until the required mode EMS or TENS and body part are selected The selected the...

Page 7: ...lace It may be helpful between applications to moisten their adhesive surface with a few drops of water be careful not to over saturate and then let them air dry to help them last longer Reusable self adhesive electrodes 10 After treatment is finished Once the treatment time has elapsed turn the intensity control dials counter clockwise to the OFF position Unplug the lead wires from the device To ...

Page 8: ...move the battery and store appropriately away from heat THERAPY PROGRAMS Program Time Rate Width Wave description TENS Burst 30 100Hz 200µs 9 pulses burst 2 bursts sec Normal 30 100Hz 200µs Continuous Mod 30 100Hz 200µs First 0 5 sec pulse width at 50 setting Next 0 5 sec pulse rate at 50 setting and width at 100 setting SD1 30 100Hz 120 Set intensity goes from 100 to 200µs 60 then back to 100 whi...

Page 9: ...the problem persists replace lead wire 3 If replacing the lead wire does not solve this problem contact us at 1 800 363 2381 Program option in use Some programs will seem intermittent This is expected Reposition and re connect the electrodes THERAPY PROGRAMS CONTINUED Program Time Rate Width Wave description EMS Neck 30 20Hz 180µs Continuous Abdomen 30 65Hz 250µs 10 second cycles 2 sec up 2 sec do...

Page 10: ...C with a relative humidity of 10 90 atmospheric pressure from 700 hpa to 1060 hpa Dimensions 110 x 65 x 30 2 mm Weight 139 g with battery Problem Possiblecause Solution STORAGE Remove the battery if the device is not in use for long periods of time Store the device in a dry place and protect it against heat sunshine and moisture Never place any heavy objects on the device DISPOSAL Used fully disch...