12
11
CLEANING AND MAINTENANCE
1. Stimulator
Remove the battery before cleaning. Wipe the stimulator with a
soft, slightly moistened cloth. If a more thorough cleaning is
needed, you can also moisten the cloth with mild soapy water.
Do not submerge the stimulator or expose it to a large amount
of water. Do not use any chemical cleaners or abrasive agents to
clean this device. After use, store the stimulator in its foam-padded
carrying case to protect it from shock and dust. If you plan not to
use it for a prolonged period, remove the battery from the unit and
store in a cool, dry place.
2. Electrodes
• Use the device only with the leads and electrodes provided
with this device or sold under the ProActive brand, following
the placements and settings prescribed by your practitioner.
• It is recommended to use at minimum 4cm x 4cm self-adhesive,
square electrodes.
• Inspect your electrodes before every use and replace as needed.
Reusing electrodes too many times may cause slight skin irritation,
low adhesion or ineffective stimulation.
• Between uses, store the electrodes in the resealed bag in a cool
dry place. It may be helpful, between applications, to moisten
their adhesive surface with a few drops of water (be careful not to
over-saturate) and then let them air-dry to help them last longer.
Reusable, self-adhesive electrodes
10. After treatment is finished
Once the treatment time has elapsed, turn the intensity
control dials counter-clockwise to the "OFF" position.
Unplug the lead wires from the device.
To remove the electrodes, lift one edge and gently peel
(do not pull on the lead wires because this may
damage the electrodes). Place the
electrodes on the liner and remove
the lead wire by twisting and
pulling at the same time.
11. To consult the memory
With the unit shut off, press and hold
M
and turn one intensity control knobs to enter
memory mode. A
M
will be displayed on
the screen.
NOTE: The maximum number of
records is 30, after which the oldest record
will be deleted.
Use the arrow keys to navigate between records and
S
to switch
between data withing one record. To clear the memory, press
the quick select button.
12. To turn off the device
Rotate both intensity controls to "0".
13. Low battery indicator
When the indicator flashes, replace the batteries as soon as
possible. However, the stimulator will continue to operate for
several hours. If stimulation seems weaker than usual, replace
the battery.