5
9.
Press the STANDBY/OPERATE switch button on the control panel (OPERATE
LED will now be illuminated and the control unit will be in operation)
Auto Set Mode
Ensure the auto sensor connector is connected properly before pressing Auto
SET button. When Auto SET function is activated, control unit is automatically
optimizing patient’s comfort setting base on patient’s BMI input. Press the
PATIENT CARE button for fast inflation. Allow 4-7 min for full inflation.
1.
Press the SET button to enter into settings mode to input patient’s BMI.
Settings is divided into three portions. The first settings mode will allow
for selection between Inch/Lbs & Cm/Kgs (Height and Weight will flash
on LCD screen during this mode). Press the SET button a second time to
enter into the next settings mode that allows for selection of height
(Height will flash on the LCD screen during this mode). Press the SET
button again to enter into the last settings mode that allows for selection
of weight (Weight will flash on the LCD screen during this mode). Finally
press the SET button to exit settings mode completely with your selected
settings for height & weight.
2.
When the mattress is fully inflated, the caregiver can transfer the patient
onto the mattress. (Note: the mattress can be inflated while a patient is
laying on it)
3.
Press PATIENT CARE again to return previous setting.
4.
By activating the PATIENT CARE function, all chambers of the mattress
system are inflated with maximum system pressure for 30 minutes. After
30 minutes, the system defaults back to previous setting.
Alternate Mode
1.
Press the "MODE" button to select the Alternate and Static Function to
enable the 3-1 alternating functions.