background image

 

 

Alarm On/Off         

 

The Alarm will be triggered when a Sensor Disconnect & Bottoming Out is 

detected.  Disable the alarm by pressing the button. 

         

 

        

Lock Button    

   

1.

 

The Prius DynaFlow is equipped with auto-locking intelligence. All 

function keys will be automatically disabled if the control panel is not in 

operation for 2 minutes and when this function is engaged an green LED 

will illuminate.  

2.

 

To unlock the control panel, simply press and hold the “LOCK” button for 

5 seconds. 
 

Comfort Level Setting 

   

 

Note:

 The pressure level settings on the weight chart are only a guideline. The 

proper adjustment of the pressure level must be applied according to 

individual patient. 

 

Please be aware that once patients have any uncomfortable feelings or 

symptoms, patients must notify healthcare professionals or caregivers to 

change product settings suit for patients‘ condition. 

 
 
 
 
 

Summary of Contents for DynaFlow

Page 1: ...g Low Air Loss Pressure Relief System User Manual Control Unit FC PHR0028 Mattress FM PHR0008 Manufactured by Prius Healthcare USA 4027 Tampa Rd Suite 3200 Oldsmar FL 34677 Tel 813 854 5464 Fax 813 854 5442 www priushcusa com ...

Page 2: ...res 3 Technical Data 3 Master Control Unit DynaFlow Mattress Replacement Symbol Definition 4 Instruction for Proper Use 4 5 Cleaning 8 The Mattress The Master Control Unit Replace Air Filter Waste Disposal 6 Storage and Care 10 7 Maintenance and Troubleshooting 10 8 EMC Related Notification 11 9 Warranty 16 ...

Page 3: ... place a heating device on or close to the mattress system Caution The Alternating System should always be used in accordance with your Institutions pressure care guidelines Re positioning of the patient is always recommended when using an alternating pressure air mattress APAM The Control unit can only be repaired by an authorized technician Do not drop the control unit Do not store the system in...

Page 4: ... or caregivers who are at least fifteen years in age with the ability to read and understand English and Westernized Arabic Numerals This device should not be operated by patient The DynaFlow Control Unit Features User friendly controls Large LCD display on each function status CPR quick release Patient Care mode provides quick maximum inflation within seconds to help transfers and nursing procedu...

Page 5: ...foam overlays or other materials which can be found in sample supply in the Prius Healthcare USA product range Therapy Mode Selection for Alternation Pulsation and Static Maximum Inflation Cycle Time Preset Key Power Switch Audio Alarm On Off Auto Set Panel Lock Power Failure Alarm Sensor Disconnect Bottom Out Alarm Comfort Setting LCD Display Patient s BMI Setting ...

Page 6: ...s Environment Temperature Operation 15 C to 35 C 59 F to 95 F Storage 5 C to 60 C 41 F to 140 F Environment Humidity Operation 30 to 75 non condensing Storage 30 to 90 non condensing Operation Atmospheric Pressure Range 700 hPa to 1060 hPa Operation altitude 1017 feet to 9 843 feet 310 meters to 3000 meters Standard IEC 60601 1 CAN CSA C22 2 No 601 1 IEC 60601 1 2 Mattress Replacement Model No FM ...

Page 7: ...nt mattress so that the air tube is at the foot of the bed Remove the mattress replacement from the box and place it directly on the bed 3 Secure straps beneath the mattress to the bed frame 4 Position the control unit on the foot board of the bed frame 5 Attach the air tube connector and auto sensor connector to control unit s socket 6 Verify that air hoses are not kinked under the mattress 7 Att...

Page 8: ...cond time to enter into the next settings mode that allows for selection of height Height will flash on the LCD screen during this mode Press the SET button again to enter into the last settings mode that allows for selection of weight Weight will flash on the LCD screen during this mode Finally press the SET button to exit settings mode completely with your selected settings for height weight 2 W...

Page 9: ...s buttocks between the cells and foam The patient should have at least 4cm of clearance between the buttocks and the bottom of the mattress If the STATIC function is selected the time display will remain blank Note The caregiver can select the Static Mode to provide the patient with only low air loss therapy To go back to Alternate Mode needs manual switch Fowler When Fowler function is activated ...

Page 10: ...hen this function is engaged an green LED will illuminate 2 To unlock the control panel simply press and hold the LOCK button for 5 seconds Comfort Level Setting Note The pressure level settings on the weight chart are only a guideline The proper adjustment of the pressure level must be applied according to individual patient Please be aware that once patients have any uncomfortable feelings or sy...

Page 11: ... can be washed and thermally disinfected in a washing machine by following below procedure Never use phenol based cleaning solutions Industrial Break washes Cold 10 minutes Main washes 60 C 140 F 6 minutes Main washes 70 C 158 F 10 minutes Extraction 2 minutes Cold Rinses Extraction 5 minutes Domestic Pre wash Cold Main Wash 70 C 158 F 10 minutes Extraction 2 minutes Cold Rinses Extraction 5 minut...

Page 12: ...back of the control unit Replace Air Filter 1 Remove air filter and replace with a new one 2 Use a soft bristle to remove dust and difficult dried on soil NOTE 1 Do not use phenol based cleaning solutions 2 Switch off the electrical supply to the pump and disconnect the power cord from the main supply before cleaning and inspection Waste Disposal This Product has been supplied from an environmenta...

Page 13: ...ooting No daily maintenance is required It is intended this equipment should only be serviced by properly qualified authorized technical personnel In case of minor trouble please refer as following Troubleshooting Symptom Inspection Procedures Possible Solution The pump is not functioning 1 Check for correct power voltage connected 2 Check for blown fuse 1 Connect to correct main power source 2 Re...

Page 14: ...nvironment RF emissions CISPR 11 Group 1 The DYNAFLOW uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The DYNAFLOW is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supp...

Page 15: ...61000 4 5 0 5kV 1kV line s to line s 0 5kV 1kV 2kV line s to earth 0 5kV 1kV line s to line s Not applicable Mains power quality should be that of a typical home healthcare environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 Voltage dips 0 UT 0 5 cycle 0 UT 1 cycle 70 UT 25 30 cycles Voltage interruptions 0 UT 250 300 cycle Voltage dips 0 ...

Page 16: ...W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE1 At 80 MHz and 800 MHz the higher frequency range...

Page 17: ...ter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 N A 0 12 0 23 0 1 N A 0 38 0 73 1 N A 1 2 2 3 10 N A 3 8 7 3 100 N A 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where p is the maximum output power ...

Page 18: ...00 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18 Hz 2 0 3 28 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217 Hz 2 0 3 28 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 2 0 3 28 28 5240 5100 5800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 9 5500 57...

Page 19: ... any equipment returned for that purpose and to replace or repair any part which is proven to be defective at no charge This warranty excludes equipment damage through shipping tampering improper maintenance careless accident negligence or misuse or products which have been altered repaired or dismantled other than with the manufacture s written authorization and by its approved procedures and by ...

Page 20: ......

Page 21: ...AL300217 V1 00 Prius Healthcare USA 4027 Tampa Rd Suite 3200 Oldsmar FL 34677 Tel 813 854 5464 Fax 813 854 5442 www priushcusa com ...

Reviews: