16.2.14 Thermoelectric Cooler
The thermoelectric (TE) cooler is mounted to the heat removal block. With air-cooling
alone, at an ambient temperature of 25
C, temperature lock at 25
will generally occur
within ten to fifteen minutes. Note that the exact cooling performance is a function of the
CCD array installed. Also, if the ambient temperature is particularly warm, achieving
temperature lock might take longer, or may not occur at all.
16.2.15 Trigger In
This is a standard BNC connector, and serves as the main trigger input to the timing
generator when external triggering is used.
When external triggering is selected in the software, the internal timing generator will be
triggered by an externally-derived trigger pulse (range of ±5.5 V) applied to this input. The
threshold, slope, coupling mode (i.e., AC or DC,) and input impedance (i.e., High or 50
,)
are selectable in software.
The adjacent green
TRIGGER IN
LED flashes each time the camera is triggered (glows
steadily at high repetition rates.) The actual triggering can also be readily determined by
observing the signal at the
Monitor
output with a fast oscilloscope.
16.2.16 User RF Out [PI-MAX4: 1024i-RF only]
This is a standard female BNC connector. User-programmable continuous frequency output
to synchronize external devices with PI-MAX4: 1024i-RF. When used, this must be
connected to a standard 50
Ω
RF load.
WARNING!
Always turn the power off at the power supply before connecting or
disconnecting any cable that interconnects the camera and the
computer or serious damage to the CCD may result. This damage is
NOT
covered by the manufacturer’s warranty.
16.3 Gig-E Ethernet Card
This card must be installed in the host computer to control the PI-MAX4
camera and to receive data from the camera.
16.4 Extender Bracket Kit
Each PI-MAX4 camera is shipped with an Extender Bracket kit that, when mounted to the
rear of the camera, allows the PI-MAX4 to be secured to a laboratory table using tapped
mounting hole arrays with ¼-20 on 1-inch spacing, or M6 on 25 mm spacing.
Summary of Contents for PI-MAX4
Page 1: ...PI MAX4 Camera System 4411 0139 Issue 9 July 3 2018 ...
Page 4: ...4 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 18: ...18 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 32: ...32 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 46: ...46 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 96: ...96 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 156: ...156 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 172: ...172 PI MAX 4 System Manual Issue 9 ...
Page 216: ...216 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 248: ...248 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 258: ...258 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 274: ...274 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 292: ...292 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 296: ...296 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 304: ...304 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 310: ...310 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 326: ...This page is intentionally blank ...