Perform the following procedure to configure a single trigger DIF acquisition:
1.
The PI-MAX4 camera must be aligned and focused on the area of interest for this
experiment. This is best accomplished while the PI-MAX4 is operating in
Full Frame
readout mode (i.e., before switching to
DIF
mode on the
Readout
expander.)
Verify that the
Phosphor Decay Delay
is appropriate to the phosphor used by the camera.
The
Phosphor Decay Delay
time entered in LightField can be viewed or changed by
clicking on the
Advanced
button on the
Common Acquisition Settings
expander.
NOTE:
The
Phosphor Decay Delay
setting tells LightField how long
to wait following the gate pulse before shifting the image. If
there is some residual image from the first frame in the second
frame, simply increase the
Phosphor Decay Delay
setting to
allow more time for the phosphor emission to decay before
shifting the image. If residual image is not an issue, then the
Phosphor Decay Delay
setting can be decreased to reduce the
time between the two DIF images.
2.
After the alignment and focus, the PI-MAX4 system needs to be placed into DIF mode.
•
On the
Readout
expander, select
DIF
as the
Mode
.
•
On the
Common Acquisitions Settings
expander, configure the
Frames to Save
to a multiple of 2 (i.e., 2, 4, 6, etc.)
3.
On the
Trigger
expander, verify that
Trigger Response
is set to
Readout Per Trigger
.
4.
On the
Trigger
expander, select
Internal
or
External
triggering.
External
: Verify the trigger characteristics on the
Trigger
expander match the active
trigger edge for the trigger pulse being used.
Internal
: Perform the following procedure to configure the
Internal Trigger
parameters
on the
SuperSYNCHRO Timing
expander:
a.
Use the hyperlinks at the top of the
SuperSYNCHRO Timing
expander to make any
changes to:
•
Intensifier Settings
;
•
Trigger Settings
;
•
Phosphor Decay Delay
;
•
Frames to Save
(must be a multiple of 2);
•
Readout Mode
(must be set to
DIF
.)
b.
Configure the
Internal Trigger Frequency
.
c.
Configure the
Gate Width
and
Gate Delay
times for each of the two images.
•
The
Initial Gate Delay
time must be
≥
86 µs.
•
The
Second Gate Delay
time must be greater than:
Phosphor Decay + Minimum Camera Delay
Typical camera values are:
—
Phosphor Decay = 1 µs
—
Minimum Delay = 30 ns
Therefore, a typical minimum value is 1.03 µs.
Summary of Contents for PI-MAX4
Page 1: ...PI MAX4 Camera System 4411 0139 Issue 9 July 3 2018 ...
Page 4: ...4 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 18: ...18 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 32: ...32 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 46: ...46 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 96: ...96 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 156: ...156 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 172: ...172 PI MAX 4 System Manual Issue 9 ...
Page 216: ...216 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 248: ...248 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 258: ...258 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 274: ...274 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 292: ...292 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 296: ...296 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 304: ...304 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 310: ...310 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 326: ...This page is intentionally blank ...