42
PI-MAX
®
4 System Manual
Issue 9
3.7
Connect the Circulator [Liquid-Cooled Cameras Only]
For liquid-cooled cameras, the CoolCUBE
II
circulator provides a vibration-free method of
heat removal.
REFERENCES:
For additional information about coolant hoses, refer to
, on
illustrates a typical CoolCUBE
II
circulator.
Figure 3-3: Typical CoolCUBE
II
Circulator
Perform the following procedure to connect a PI-MAX4 to a CoolCUBE
II
circulator:
1.
Make sure the camera and the circulator power switches are turned off.
2.
Make sure the circulator is 6 inches (150 mm) or more below the camera. The vertical
distance should not exceed 10 feet (3 m.) Typically, the camera is at table height and the
circulator is on the floor.
3.
Make the coolant connections between the circulator and the camera. It does not matter
which hose from the circulator is plugged into a coolant port on the camera.
4.
It is recommended that hoses be secured to the camera hose barbs with the clamp
supplied.
NOTES:
1.
Verify that there are no kinks in the hoses that impede the
coolant flow. Lack of sufficient flow can seriously harm
the camera and any resulting damage is not covered under
warranty.
2.
Damage caused by water leaking into the camera voids the
warranty.
5.
Unscrew the reservoir cap (on top of the CoolCUBE
II
) and make sure that the coolant
reservoir contains coolant. If additional coolant is required, fill with a 50:50 mixture of
water and ethylene glycol.
6.
Screw the reservoir cap back in.
Coolant Ports
Reservoir Cap
4411-0139_0007
Summary of Contents for PI-MAX4
Page 1: ...PI MAX4 Camera System 4411 0139 Issue 9 July 3 2018 ...
Page 4: ...4 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 18: ...18 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 32: ...32 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 46: ...46 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 96: ...96 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 156: ...156 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 172: ...172 PI MAX 4 System Manual Issue 9 ...
Page 216: ...216 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 248: ...248 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 258: ...258 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 274: ...274 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 292: ...292 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 296: ...296 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 304: ...304 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 310: ...310 PI MAX 4 System Manual Issue 9 This page is intentionally blank ...
Page 326: ...This page is intentionally blank ...