Cube L1
49
Revision 03 EN - 03/2021
16.1.2 Exchanging the protective window
1.
Observe the safety instructions in chapter 16.1.1, „Safety instructions“, on
page 47.
2.
Unscrew the 3 Torx screws M3 x 6 mm on the protective window holder.
3.
Place the device as shown in Fig. 16.1 on page 49 and carefully remove the
protective window holder upwards.
•
Make sure that the inserted O-ring does not fall out of the device.
4.
Remove the old protective window from the device and dispose of it.
5.
Wear powder-free latex gloves and insert the new protective window into the
device.
•
Ensure that the inserted O-ring is not out of place.
6.
Place the protective window holder according to Fig. 16.1 on page 49.
7.
Tighten the protective window holder with 3 Torx screws M3 x 6 mm.
8.
Check for secure fit of the protective window holder:
•
The protective window holder must lie flat on the device.
Torx screws
M3 x 6 mm
Protective window
holder
Optics
O-ring
Protective
window
Fig. 16.1: Exchanging the protective window on the Cube L1