35,0(6
Cube L1
39
Revision 03 EN - 03/2021
14.4
Settings for trigger level, power saving functions and automatic
measurement readiness
The trigger level determines the sensitivity of the photodiode for laser detection. The
trigger level Low is recommended for pulsed beams.
With a low trigger level a measurement may be triggered even without a laser beam.
In this case set the trigger level to a higher value. The default setting for the Cube L1 is
trigger level Medium.
1. Set the
Trigger level (Laser
power)
.
•
In the Cube L1 the default set-
ting is the Trigger level
Medium
(
100W-1000W
).
2. Click on the
Save settings
button.
In the LaserDiagnosticsSoftware LDS
you can make additional settings for
power saving functions and the auto-
matic measurement readiness:
1. Enter the desired settings in the
Device control
.
2. Click on the
Save settings
button.
Autom. ready for measurement
By default, the device automatically returns to
measurement readiness after each measurement. If
you uncheck the box, you will need to reset the unit
after each measurement by briefly pressing the on/of
button.