BeamMonitor BM+
35,0(6
35
Revision 02/2019 EN
17.3
Replace the detector
Generally, the BeamMonitor BM+ is equipped with a DBIG-PS+ or DBC+ detector. Detectors with different
sensitivity or different time behavior can be used for special applications (siehe Tab. 17.1 on page 34).
17.3.1
Remove cover
1. Turn off the power supply.
2. Unscrew four torx screws T8 from the cover.
3. Lift the cover to remove it.
• Under the cover is the detector.
Torx T8
Torx T8
Torx T8
Torx T8
Fig. 17.1:
Opened cover of BeamMonitor BM+ with detector
17.3.2
Disassemble the detector
NOTICE
Damage of the detector
Touching the sensor surface will damage the detector. This can negatively affect the measur-
ing results.
X
Do not touch the detector with your fingers on the sensor surface.
X
Don’t set the detector on the sensor surface.
Sensor surface
Fig. 17.2:
Sensor surface on the detector