BeamMonitor BM+
35,0(6
27
Revision 02/2019 EN
10.3.3
Configuring advanced settings (
Device control > Advanced
)
1. Click on the
Advanced
tab.
Resolution of the measurement window
2. Enter the number of
Pixel in x/y
for
the measurement.
Wave length
3. Select the
Calibrated wavelength in
nm
of your BeamMonitor BM+.
4. Enter the
Used wavelength in nm
of
the laser.
Laser power and focal length (option)
The input of the laser power and focal
length is needed for the calculation of the
beam quality factor M², the power density
and the raw beam diameter, etc.
5. Enter the laser
Power P in W
.
6. For a built-in focusing optics, enter
the
Focal length of focusing optic
in mm
.
Averaging
By placing a check mark next to
Averag-
ing
, you can select various algorithms for
averaging the measuring values:
Analysis of series measurements. Averag-
ing algorithms: Average value, value of the
maximum pixel and value of the maximum
trace.
7. Enter the
Number of averaged
planes
.