![Primes BeamMonitor BM+ Series Operating Manual Download Page 10](http://html1.mh-extra.com/html/primes/beammonitor-bm-series/beammonitor-bm-series_operating-manual_1601730010.webp)
BeamMonitor BM+
35,0(6
10
Revision 02/2019 EN
3
About this operating manual
This documentation describes the installation and configuration of the BeamMonitor BM+ and the execution
of measurement with the PRIMES LaserDiagnosticsSoftware LDS.
The LaserDiagnosticsSoftware LDS must be installed on the PC for measuring operation of the BeamMonitor
BM+. The basic version of the LaserDiagnosticsSoftware LDS is included in the scope of delivery for the
device.
For a detailed description of the software installation, file management and evaluation of the measured data,
please refer to the separate operating manual LaserDiagnosticsSoftware LDS.
4
Conditions at the installation site
• The device must not be operated in a condensing atmosphere.
• The ambient air must be free of organic gases.
• Protect the device from splashes of water and dust.
• Operate the device in closed rooms only.
DANGER
Fire and explosion hazards due to scattered or directed laser radiation
When the BeamMonitor BM+ is being operated, the irradiation must be fully absorbed be-
hind the measurement zone. Fire bricks or other partly-absorbing surfaces are not suitable.
X
Use an adequate absorber. Dependent on the application, PRIMES offers suitable absorbers,
such as the PowerMonitor PM 48/100.
X
Don’t store any flammable materials or highly flammable substances at the measuring loca-
tion.