Page 8
- Pressure Inputs
Ranges *
Resolution Accuracy**
Remarks
(0) 0
– 250 mmH
2
O
0.001
0.05 %
Gage pressure.
Used with air or
inert gases.
(1) 0
– 1 psi
0.0001
0.05 %
(2) 0
– 5 psi
0.0001
0.025 %
(3) 0
– 15 psi
0.0001
0.025 %
Gage or absolute
pressure.
Used with fluids
(gases or liquids)
compatible with
316 L stainless steel.
(4) 0
– 30 psi
0.0001
0.025 %
(5) 0
– 100 psi
0.001
0.025 %
(6) 0
– 250 psi
0.001
0.025 %
(7) 0
– 500 psi
0.01
0.025 %
(8) 0
– 1000 psi
0.01
0.025 %
(9) 0
– 3000 psi
0.01
0.025 %
(10) 0
– 5000 psi
0.1
0.025 %
(11) 0
– 10000 psi
0.1
0.05 %
(12) Other, upon request
(*) Gage pressure, vacuum (only for range 3), differential (from range 0 to 2), absolute or compound (from range 3 to 8).
(**) Full Scale percentage.
Accuracy values are valid within one year and ambient temperature of 20 to
26 °C. Outside these limits add 0.005 % FS / °C, using 23 °C as the reference
temperature.
presys
Summary of Contents for PC-507-IS
Page 1: ...PC 507 IS Pressure Calibrator TECHNICAL MANUAL presys Instruments p r e s y s ...
Page 10: ...Page 9 2 3 Order Code p r e s y s ...
Page 13: ...Page 12 3 Operation 3 1 Parts Identification Front Panel Fig 01 Front Panel p r e s y s ...
Page 14: ...Page 13 Left side view Panel Right side view Panel Fig 02 Side Panels p r e s y s ...
Page 31: ...Page 30 a 4 wire Transmitter Calibration Fig 13 4 WIRE Transmitter Calibration p r e s y s ...