Page 16
3.3. Using the Calibrator: Basic Functions
When the calibrator is powered on, the display shows:
OFF
ON
If the user does not select ON and press ENTER within 12 seconds, the
instrument will turn itself off. The purpose is to save battery from turning the
instrument on accidentally while in the bag.
When
ON
is confirmed, the calibrator goes through a self-test routine and
shows the last adjustment date and the value of the battery voltage; in case of failure,
it displays a message to indicate RAM error or E2PROM error; if that occurs, the
instrument should be sent for repair.
After the self-test, the display shows the starting menu:
IN
OUT
EXEC
CONF
ADJ
COM
IN / OUT
- selects the input/ output functions.
ADJ
- selects functions for the adjustment of the calibrator itself (See adjustment
section). Do not enter in
ADJ
option before reading the warning in section
4 - Adjustment.