
Page 52
b) Standard Resistors Application:
Connect a decade box or standard resistors to the
Probe
connector, at
positions RTD1, RTD2, RTD3 and RTD4, as shown above.
Resistors
1
st
point
2
nd
point
R_2
20.000
50.000
R_1
100.000
500.000
Pressure Adjustment
By selecting
Press
(ENTER), it shows the sub-menu:
V_1
V_2
P1
P2
Pressure
input adjustment is only completed after calibrating the options:
V_1
,
V_2
,
P1
and
P2
. Note that
P1
and
P2
refer to two pressure sensors chosen
from
P
option in
CONF
menu
The
V_1
and
V_2
adjustment is performed only at the factory.
The calibrations of the sensors associated to mnemonics
P1
and
P2
consist of
the application of two pressure values close to zero and full scale (FS) of the
pressure sensor. Store the values of the 1
st
and 2
nd
calibration points by means of
keys 1 (1
st
point) and 2 (2
nd
point).
presys
Summary of Contents for PC-507-IS
Page 1: ...PC 507 IS Pressure Calibrator TECHNICAL MANUAL presys Instruments p r e s y s ...
Page 10: ...Page 9 2 3 Order Code p r e s y s ...
Page 13: ...Page 12 3 Operation 3 1 Parts Identification Front Panel Fig 01 Front Panel p r e s y s ...
Page 14: ...Page 13 Left side view Panel Right side view Panel Fig 02 Side Panels p r e s y s ...
Page 31: ...Page 30 a 4 wire Transmitter Calibration Fig 13 4 WIRE Transmitter Calibration p r e s y s ...