Page 40
Fig. 21
– Function SCALE (LINEAR)
The number of decimals (up to 4) shown on the display may be configured by
using
Scale Dec
parameter.
The value for
Input High
must be necessarily higher than
Input Low
. On the
other hand,
Scale High
and
Scale Low
may have any relationship between
themselves: higher than, lower than or equal to, and they may have a signal before
them. Thus direct or reverse relationships may be established.
For the current input, a linear relationship may be established as it has been
previously shown or it may be squared
(FLOW)
as illustrated below:
presys