Page 33
f) 2-wire Pressure Transmitter Calibration.
Use the 12V@20mA power supply from the calibrator (
TPS
) to power a 2-wire
transmitter and make the connection of current as follows.
Select
mA
(current) from the input menu of the calibrator and
Pressure
from
the output menu. Input current readings can be scaled to pressure unit through the
special function
CAL
(see item about Special Functions). In this way, the deviation
between the input and the output of the pressure transmitter can be easily calculated.
Fig. 18
– 2-wire Pressure Transmitter Calibration
presys