7 Special operating keys
■
43
7
Special operating keys
7.1
The Tare key «0/T»
•
Effects taring
– Ensure that the balance is in the weighing mode.
– Briefly press «
0/T
».
– The balance performs a tare operation.
•
Effects a calibration
– Ensure that the balance is in the weighing mode.
– Keep
«
0/T
»
pressed until „CAL“ is displayed.
– Release «
0/T
».
– The balance carries out a calibration in accordance with the
settings in the configuration menu (see chapter 5.5 “Calibration
functions”) and reports these by means of a printout..
NOTE
A calibration and/or a calibration-test can be cancelled with
«
ON/OFF
».