14 Further information
62
14.2.2 External calibration with freely selectable weight
For an external calibration with user-definable weight, „ModE DEF”
must be selected in the configuration menu (see chapter 5.5 “Calibra-
tion functions”).
Then, the effective value of the calibration weight (DEF. n.nnn g) must
be entered with up to tenfold precision compared with the balance.
Then proceed as follows:
NOTE
If calibration is carried out with the free weight, then only this
weight may be used.
• Switch to „bALAncE” with the
«MODE»
-key.
CAL
• Press «
T
» until „CAL” appears.
- - 0000
9
• The balance carries out a Zero measurement
(- - 0000 g is shown flashing).
- - 232
9
• After the zero measurement the display
flashes with the previously entered calibration
weight.
• Place the calibration weight on the pan.
- - 232
9
• The display flashes rapidly.
• Calibration is complete when the display stops
flashing (the exact weight is shown).
+ 232
9