WMA-5 Operation Manual V. 1.01
19
2. The units associated with the parameter.
3. The acceptable range of values.
4. The value being entered.
5. The OK button selects the entered value. If the entered value is out of range, it is set to the
minimum or maximum value. It then returns to the previous screen.
6. The Cancel button returns to previous screen without changing the initial value.
Measurement Mode
The Measurement Mode is comprised of four screens; Measure screen, Variables 1 screen, Variables 2
screen, and the Graphic Display screen. Each screen allows the user to view the WMA-5 data in various
ways. The screens can be rotated through by the user using the left and right arrow buttons. The initial
screen is the Measure screen which displays the value of CO
2
concentration in a large bold font. The
Second screen is the Variables 1 screen which displays the values of six variables in real time; CO
2
, H
2
O,
O
2
, Temperature, Flow Rate, and Pressure. The Third screen is the Variables 2 screen which displays
the values of four variables in real time; CO
2
, H
2
O, O
2
, and Aux. The forth screen is the Graphic Display
screen which displays CO
2
, H
2
O or O
2
concentration graphically over time.
Measure Screen Info
Each measurement screen contains these common features.
1. There is a pulsing
♥
icon to confirm that the WMA-5 display is actively on.
2. The Record button saves data as a marked record in the USB memory stick, and also sends the
record to the host and WiFi ports.
3. The percentage value in the top right corner is the estimated
percentage of Auto-zero absorber
column capacity remaining based on the number of zeros performed since the last time the
counter was reset by the user (See Reset Zero Absorber on page xx).
4. The ‘Z’ button is used to initiate a manual zero.
5. A status or error message can be displayed in the Status Area.
1
3
4
5
2