WMA-5 Operation Manual V. 1.01
54
during this time). This command can be initiated by the user or by the WMA-5 itself.
Example:
Sent from Host: Z<CR>
Received from WMA-5:
+<CR>
Z, 025<CR>
Z, 024<CR>
…
A zero has been initiated by the user. In this example, xxx is 025 then 024 which
counts down of the number of cycles remaining.
Get/Set Parameter Values Table
Get/Set Parameter Values
Number
(X)
Parameter
Name
Description
Valid
Values
Default Value
1
Zero Type
See Zero Settings section. The
values are: 1=Automatic, 2=User
Set, 3=Manual
1-3
1 (Automatic)
2
Zero Time (min)
See Zero Settings section.
1-40
20
3
Average Limit
(ppm)
See Averaging Settings section.
0-100
30
4
Alarm Low (ppm)
See Alarm Settings section.
0-1000
250
5
Alarm High (ppm)
See Alarm Settings section.
100-20000
2000
6
Alarm Sound
See Alarm Sound section.
0=Off;
1=On
0 (Off)
7
CO
2
Concentration
(ppm)
See Analog Output Settings
section.
100-
100000
2000
8
WMA ID
See WMA ID section.
0-9
1
9
Pump Power (%)
Percentage of the pump power (i.e.,
air flow).
0-100
60%
10
Host Measure
Format
See Host Measure Settings section. 1-6
3 (Full format)
11
Host Measure
Interval (seconds)
See Host Measure Settings section.
Use ‘0’ for manual mode to stop
broadcast.
0-3600
1 second
12
Memory Measure
Format
See Memory Format Settings
section.
1-6
3 (Full format)
13
Memory Measure
Interval (seconds)
See Memory Measure Settings
section.
0-3600
1 second
14
WIFI Measure
Format
See WIFI Measure Settings section
1-6
2 (No header
format)
15
WIFI Measure
Interval (seconds)
See WIFI Measure Settings section. 0-3600
0 seconds (Manual)
16
Not Used
17
CRC
Determines whether a CRC-8 will
be appended to each command
received from the WMA-5. This
works for the Host only.
0=Off;
1=On
0 (Off)
18
Web Monitor
Adds counter and terminator to
strings sent to the WiFi Port to use
for web pages
0=Off;
1=On
1 (On)