WMA-5 Operation Manual V. 1.01
53
Example:
Current time: T,04,42,15,05,01,15<CR>
Sent from Host: T,04,42,10,05,02,15<CR>
Received from WMA-5:
+<CR>
T,04,42,10,05,02,15<CR>
In this example, hour is changed from 15 to 10 and month is changed from 1 to 2.
If any one of the attributes is different from the current the value, it will be updated.
3. A Partial Set: The ‘T’ command can be used to set particular attributes of the
day/time values using 1-6, where 1=seconds, 2=minutes, 3=hour, 4=day, 5=month
and 6=last 2 digits of year.
Example:
Current time: T,4,42,10,05,02,15<CR>
Sent from Host: T,3,11<CR>
Received from WMA-5:
+<CR>
T,4,42,11,05,02,15<CR>
In this example, the hour was changed from 10 to 11. All other values stayed the
same.
V
Standard This command returns the serial numbers and software versions of the WMA-5
controller, the IRGA and the Display.
Example:
Sent from Host: V<CR>
Received from WMA-5:
+<CR>
V,WMA500007,01.02,IRGA00001,01.02<CR>
Where, in this example,
WMA-5 serial number= WMA500007
WMA-5 version number=01.02
IRGA serial number=IRGA00001
IRGA version number=01.02
W
Auto
During warm up of the WMA-5, the W command is transmitted every second until the
operating temperature is reached. (No “M” strings are sent during this time)
Example:
Received from WMA-5:
W,55.0<CR>
Where in this example, 55.0 is the IRGA temperature (°C).
X
Standard This command retrieves last seven zeros performed.
Example:
Sent from Host: X<CR>
Received from WMA-5:
+<CR>
X, 44200, 44223, 49971, 49972, 49976, 49981, 49973<CR>
The last seven zeros are displayed.
Z
Z,xxx
Auto &
Standard
This command initiates a zero. During a zero operation, the WMA-5 transmits every 1
seconds the Z command until the end of the zero operation. (No “M” strings are sent