WMA-5 Operation Manual V. 1.01
8
Technical Specification
Analysis Method
Non-dispersive infrared, configured as an absolute absorptiometer with
microprocessor control of linearization
CO
2
Measurement
Ranges
0-1000 ppm (µmol mol
-1
)
0-2000 ppm (µmol mol
-1
)
0-5000 ppm (µmol mol
-1
)
0-10000 ppm (µmol mol
-1
)
0-20000 ppm (µmol mol
-1
)
0-30000 ppm (µmol mol
-1
)
0-50000 ppm (µmol mol
-1
)
0-100000 ppm (µmol mol
-1
)
Readings are automatically corrected for temperature and pressure.
Accuracy
< 1% of span concentration over the calibrated range but limited by the
accuracy of the calibration mixture.
Linearity
< 1% throughout the range
Stability
Auto-zero at regular intervals corrects for sample cell contamination,
source and detector ageing and changes in electronics.
Warm-up Time
Approximately 15 minutes
Sampling Rate
10 Hz. Sample data is averaged and output every 1.0 seconds.
Sampling Pump
Integral, long life 12V air sampling pump
Gas Flow Rate
200-500 cc/min (280-340 cc/min is optimal). An internal, electronic flow
sensor monitors flow rate.
Terminal Block
System power, analog, digital and current (4-20mA) outputs and sensor
inputs are available from the 22 pin terminal block
Analog Output
0-2.5V (CO
2
range selectable) and 4-20mA
Digital Output
RS232 and USB
Air Filter & Water Trap
An external water trap and internal hydrophobic filter are used to protect
analyzer from water.
Alarm
Visual and audible alarm/warnings. 2 relay contacts (Alarm1 and Alarm2)
CO
2
Control
High and low user set points
Data Storage (USB)
USB Flash Drive for data storage in multiple formats.
Display
2.7” electronic paper touch display with 264 x 176 pixel resolution
Power Requirements
12 VDC, 1.5A or
100-240 VAC, 50-60 Hz, 0.6A (AC Adapter included)
Power Consumption
Warm up: 12W (12V @ 1.0A)
Normal operation: 6W (12V @ 0.5A)
Enclosure
High impact, IP65 enclosure
Gas Connections
Two barbed fittings (inlet and exhaust) for use with 1/8” (.125”) ID tubing
Operating Temperature
0-50
o
C, non-condensing. External filtration may be required in dirty
environments.
Dimensions
213 cm L x 185 cm H 114 cm W (Enclosure only)
Weight
1.5 kg
Optional Accessories
-
H
2
O Sensor - Range: 0-Dewpoint (mb)
-
O
2
Sensor – Range: 0-100%
-
WIFi
PP Systems is a registered trademark of PP Systems, Inc.
PP Systems is continuously updating its products and reserves the right to amend product
specifications without notice. All brand names are trademarks or registered trademarks of their
respective owners.