After having reset the factory default parameters, ACTEON 2050 will be configured with the following
parameters:
Sensor response averaging
30 measurements
SS sensor offset
0mV
Time base for trend line display
1 minute
4-20mA output adjustment
SS output: The 4mA stop is set to 0g/l and the 20mA stop is
set to 20g/l (ACTEON 2050-20) or 50g/l (ACTEON 2050-
50)
Temp. °C output: The 4mA stop is set to -5°C and the
20mA stop is set to 35°C.
Relay output adjustment
The relays are configured in mode 1:
- R1: threshold: 0.00g/l hyst: 0.20 below the threshold,
relay R1 is active.
- R2: threshold: 50.00°C; hysteresis: 0.10 above the
threshold, relay R2 is active
Temperature measurement unit
°C
Language
English
10 Information menu
The INFO window provides information on the device type and version. The software and hardware versions
will be requested when you call PONSEL Technical support or After-sales service.
INFO
ACTEON 2050 SS meter
Temperature
Software version XX.XX
Hardware version: XX.XX
ESC: To escape menu